TGS2Описание
Grating set TGS2 consists of 6 calibration gratings: TGZ1, TGZ2, TGZ3, TGT1, TGG1, TGX1.
Grating set TGS2 can be used for:
TGZ1 - for Z-axis calibration (step height 20 ± 2 nm) and nonlinearity measurements TGZ2 - for Z-axis calibration (step height 110 ± 10 nm) and nonlinearity measurements TGZ3 - for Z-axis calibration (step height 520 ± 20 nm) and nonlinearity measurements TGX1 - for lateral scanner calibration, detection of lateral non-linearity, hysteresis, creep, and cross-coupling effects, determination of the tip sharpness. TGG1 - for AFM calibration in X or Y axis, detection of lateral and vertical scanner nonlinearity, tip sharpness characterization. TGT1 - for tip shape and sharpness estimation, for image deconvolution and contamination control. |